If any troubles should occur, note the error
code, stop the operation immediately. Turn
the power switch off, pull the power cord
out and contact the seller or Yamato
Scientific’s Technical Service Department.
(Necessary information)
●Model number
●Serial number
●Date of purchase
●See the warranty or nameplate on the unit
●Detailed information on difficulties
●The service man requests you to show the
warranty on his visit.
Minimum period to keep repair parts is 7 years limited after
we stop the production of this model of units. The repair
parts mean any necessary parts to maintain the
performance of the unit.
When you encounter the following
Fault indication
Check points
STAND BY switch is not lit when the
breaker is turned on.
● Is the power supply cord is securely connected with the outlet?
● Is there any power failure?
No water is supplied.
● Defect of solenoid valve to collect ion-exchanged or distilled water
● Defect of solenoid valve to supply raw water.
● Low pressure or interruption of water service.
● Defect of pressure switch.
Water supply can not be stopped.
● Defect of float switch.
● Defect of solenoid valve to supply raw water.
No water is supplied into the boiler.
● Defect of solenoid valve to supply boiler water
● Defect of float switch.
● Stoppage in the pre-treatment cartridge.
Heater can not be turned on.
● Defect of float switch.
● Heater disconnection.
No cooling water flows.
● Defect of solenoid valve for cooling water.
Initial distilled water is not drained.
● Defect of solenoid valve to drain initial distilled water.
Distilled water is not stored.
● Defect of solenoid valve to drain initial distilled water.
● Defect in piping.
Distillation would not stop.
● Defect of float switch.
No water is collected.
● Defect of solenoid valve to collect ion-exchanged or distilled water.
● Defect of pump to take distilled water.
● Defect in piping.