Maintenance, Cleaning still – Yamato Scientific WG201 Water Purifiers User Manual
Page 31
-29-
Maintenance
Cleaning Still
Removal the still
1. Turn “OFF” the circuit breaker.
2. Close the faucet.
3. Make sure if the boiler cools down enough (about more than 30 min. later). Open the front door of this
unit and the valve to drain boiler water.
4. Remove both hose connected to the
① boiler and ② condenser respectively. Twist the hose-band by
using tools, move the part of notches and you can remove the exit of distilled water and intake/drain of
boiler. When you remove them, be careful not to damage the glasses with giving excessive power.
Remove the nut of entrance and exit of cooling water by rotating counterclockwise. Be sure not to lose
the inside packing.
5. Take off the four connecting screws by cross screwdriver, and take off the left side board.
6. By cross screwdriver, loosen the four connecting screws in right side of terminal rack that is in the upper
right of framework made by removing the left side board. Pull the heater terminal.
- BB300 Low Constant Temperature Water Bath (64 pages)
- BBL100 Low Constant Temperature Water Bath (61 pages)
- BH200 Thermo-Elite (37 pages)
- BH400 Thermo-Elite (42 pages)
- BM100 Water Baths (16 pages)
- BM110 Water Baths (16 pages)
- BM400 Water Baths/Oil Baths (22 pages)
- BM410 Water Baths (21 pages)
- BU200 Low Constant Temperature Water Bath (18 pages)
- CLH301 Cool Line (65 pages)
- CLS301 Cool Line (55 pages)
- DC400 Freeze Dryers (33 pages)
- DG400 Drying Ovens for Laboratory Tools (41 pages)
- DKN 302 Constant Temperature Drying Oven (50 pages)
- DN410H Constant Temperature Ovens (88 pages)
- DN410I Inert Oven (88 pages)
- DNF 400 Constant Temperature Ovens (76 pages)
- DVS402 Constant Temperature Drying Oven (47 pages)
- DX302 Constant Temperature Drying Oven (40 pages)
- DX302 Constant Temperature Drying Oven (9 pages)
- HC200 Cool Block (58 pages)
- HF100 Heating Block (29 pages)
- IJ100 Low Temperature Incubator (27 pages)
- IJ201 Low Temperature Incubator (50 pages)
- K400 Shaking incubator (36 pages)
- IN603 Low Temperature Incubators (68 pages)
- IN603W Low Temperature Incubators (67 pages)
- IQ820 Low Temperature Incubators (46 pages)
- IQ821 Low Temperature Incubators (47 pages)
- LT400A Laboratory Stirrer (21 pages)
- RE200 Evaporators, Rotary (29 pages)
- RE400 Evaporators, Rotary (21 pages)
- RE210 Evaporators, Rotary (28 pages)
- RE440 Evaporators, Rotary (25 pages)
- RE550 Evaporators, Rotary (22 pages)
- RT200 Solvent recovery unit (15 pages)
- SN 200 Autoclave, Sterilizers (66 pages)
- Vacuum controller Vacuum controller (34 pages)
- WG221 Water Purifiers (51 pages)
- WR600A AutoPure (28 pages)
- WS200 Water Purifiers (32 pages)
- External output controller type Ⅳ CR (3 pages)
- Hitech controller type Ⅳ (34 pages)