beautypg.com

Identification of parts, Operation system – Yamato Scientific WG201 Water Purifiers User Manual

Page 21

background image

-19-

Identification of Parts

Operation System

(1) Supply and distillation of boiler water

●Turn the circuit breaker on and push the power switch ,then,

 solenoid valve for raw water and

solenoid valve for boiler water opens at the same time ,and then water is supplied into

boiler.

When

float switch to control the heater in the

float receptacle finds water level sufficient,

heater becomes on to start distillation.

Following water supply, when

float switch controls water supply into the boiler by opening and

closing

solenoid valve for raw water and

solenoid valve for boiler water.

(2)Flow of cooling water

●In distillation of boiler water ,cooling water flows and drains from the

decompressor,

cooling

water solenoid valve to

condenser. When distilled water tank is full of water or taking ion-

exchanged water , distillation stops and water flow and drain stops automatically.