Handling Precautions

WARNING!

Substances that cannot be used

Never use explosive substances, flammable substances and substances that include explosive
or flammable ingredients in this unit. Explosion or fire may occur. (Refer to page 12 "List of
Dangerous Substances".)

If a problem occurs

If an abnormality such as strong odor of the solvent is found, please request a checkout of
equipment from selling office or sales office of Yamato Scientific. Co., Ltd. The fire disaster or
bombing may occur if the abnormality is left. Do not try to repair the equipment by yourselves
for your own safety.

Do not disassemble or modify this unit

Do not disassemble or modify this unit. Fire or electrical shock or failure may be caused.


Maintenance Method

Daily Inspection and Maintenance


For the safety use of this unit, please perform the daily inspection and maintenance without fail.

CAUTION!

• Wipe the dirt with soft cloth wrung out with mild detergent. Do not use benzene, thinner or

cleanser, or do not scrub it with a scrubbing brush. Deformation, deterioration or discoloration
may result in.