Before using this unit, Warning, Caution – Yamato Scientific RT200 Solvent recovery unit User Manual
Page 6: Requirements for installation, Modification
4
Before Using This Unit
Requirements for Installation
WARNING!
1. Do not use this unit in an area where there is flammable or explosive gas
/ Install exhauster and ventilator
• Never use this unit in an area
where there is flammable or
explosive gas.
• Place an exhauster, ventilator and
fire extinguisher near the
equipment because the toxic gas
may generat
2. Do not modify
• Modification of this unit is strictly prohibited.
This could cause a failure.
CAUTION!
3. Installation on horizontal surface
• Place this unit as flat a place as possible. If the
four rubber feet are not in uniform contact with
the floor surface, noise or vibration may result.
Additionally, the unit may cause a problem or
malfunction.
4. Before/after installing
• It may cause injure to a person if this unit falls down or moves by the earthquake and the
impact. etc..To prevent, take measures that the unit cannot fall down, and not install to busy
place.
Extinguisher
Exhauster and ventilator
Modification