6

Installation


This equipment is used to collect the solvent in the rotary evaporator.

Specifications for built-in condenser and flask

Condenser: Hard glass, Surface area of condenser
Flask: Hard glass, 500ml round flask

Installation method

The normal installation method is shown below.

A cooling water circulation system and vacuum pump are required in use in addition to the RE main
unit.
Connect the vacuum hose between the exhaust nozzle of vacuum pump and IN nipple on the
RT200, as shown in the figure. The end connection of exhaust piping and cooling water piping on
the RT200 model are placed on the back surface of main unit.

Connected to the nozzle

on the back surface of RT200

IN

OUT

真空ポンプ

IN

OUT

To ventilation system

(such as a draft chamber)

Cooling water

circulation system

or tap water

Exhaust IN

Exhaust OUT

Cooling OUT

Cooling IN

Vacuum pump

Location of nipple connecting with the hose

Cooling IN

Cooling OUT

Vacuum exhaust IN

Vacuum exhaust OUT