beautypg.com

Before using this unit, Warning, Requirements for installation – Yamato Scientific DC800 Freeze Dryers User Manual

Page 9

background image

5

Before Using this unit

Requirements for Installation

WARNING!

1. Always ground this unit

• Connect the power plug to a receptacle with grounding connectors.
• Do not forget to ground this unit, to protect you and the unit from electrical shock in case of

power surge. Choose a receptacle with grounding connectors as often as possible.

• Do not connect the grounding wire to a gas pipe, or by means of a lightning rod or telephone

line. A fire or electrical shock will occur.

• Do not use a branching receptacle, which may cause the heat generation.

2. Choose a proper place for installation

• Do not install this unit in a place where:

♦ Rough or dirty surface.

♦ Flammable gas or corrosive gas is generated.

♦ Ambient temperature bellow 5℃ or above 30°C.
♦ Ambient temperature fluctuates violently.

♦ There is direct sunlight.

♦ There is excessive humidity and dust.
♦ There is a constant vibration.

♦ Winds from the air conditioner, etc. hit the sample container directly.

• Install this unit on a stable place with the space as shown below.

More than

20cm

More than

20cm

Front side

More than 40cm

3. Do not use this unit in an area where there is flammable or explosive gas

• Never use this unit in an area where

there is flammable or explosive gas.
This unit is not explosion-proof. An arc
may be generated when the power
switch is turned ON or OFF, and
fire/explosion may result.

• To know about flammable or explosive

gas refer to page28 “List of Dangerous
Substances”.

可燃性ガス

爆発性ガス

Main

Unit