Operation method, Procedure of operation – Yamato Scientific DC800 Freeze Dryers User Manual

Page 16

background image

12

Operation Method

Procedure of Operation

Before operation, prepare a drying chamber, a vacuum pump, a vacuum hose, a tandem tube, etc.

Vacuum Pump: PD52 (YAMATO Product)

Exhaust amount: 50 ・/min, Ultimate pressure: 6.7×10

-2

Pa,

Diameter of inlet pipe: 18mm (or equivalent)

1. Supply the power

ON

• Turn "ON" the earth leakage breaker.

2. Operation of the refrigerator

DEFROST

ON

OFF

ON

OFF

REFRIGE

• Turn "ON" the refrigerator switch. The refrigerator

activates.

™ There is the case that the starting sound of the refrigerator

might become higher depending on the operation initial status
and ambient temperature. However, this phenomenon is not
abnormal.

3. Installation of chamber

1. Make sure that there is no dust adhered to the trap packing

or the chamber for flask installation.

™ If any dent or dust is existed, degree of vacuum in the trap

bath becomes low.

2. Gently put the chamber for flask installation etc. on trap

packing.

™ If it has not stuck to trap packing enough, it will become the

cause of a vacuum leak.

™ The same is said of the case of a manifold etc.

Chamber for flask
installation
(Optional)

Trap packing

Manifold A