8

Before Using this unit

Installation Method

1. Unlock the stopper of the caster.

Pulling up the lever of the stopper for
caster releases the lock.
(Only the two casters in front of the unit
are attached the stopped.)

2. Move the device to the place to be

installed.

™ If there is a step on the floor, the too strong

impact is given to the caster, and could give
the damage. In that case, move the
device by lifting at the step.

3. When the installation place is determined,

pull down the lever of the stopped for
caster, and lock them.

4. Drain Tap Check
• Check whether the drain tap is detached

or not.

5. Power Plug Connection

Check the power of the earth leakage breaker is turned "OFF", and plug the power cord in the
receptacle.

6. Prepare the vacuum pump that the exhaust amount is 50 ・/min or more, and the ultimate pressure is

1.0X10

-1

Pa. This pump must also have a check-valve.

7. Install the vacuum valve to all of the valve mounting holes on the chamber.

8. Prepare a spare freeze dryer that can retain temperature sufficiently lower than the eutectic (crystal)

point of the sample.

Caster

Stopper

Release lock

Lock

Drain tap