Operational procedure, Pre-setting temperature – Yamato Scientific BO600 Water Baths/Oil Baths User Manual
Page 15

13
Operational Procedure
Pre-setting Temperature
How to select the preset temperature.
You can select the preset temperature when setting your desirable temperature. Operate according to
the following procedure. The unit continues to work when you select the preset temperature.
Display after operation procedure
Explanation
1
Push “SET” key
ex.
The temp. last set time
was 40
℃.
Push “SET” key to select the temp.
setting mode.
→The display changes from the
current measured temp. to the
last set temp. blinking.
2
Push both
▽and △key.
Push both
▽and △key
simultaneously for more than 1
second.
→The unit changes into the
presetting mode, and the display
blinks the preset temp.
When no temp. is preset, the display
blinks
.
3
Push “SET” key.
ex.
The current measured
temp. is 30
℃.
Push “SET” key after you select the
presetting temp.
→The unit begins operation up to
the preset temp.
→The display stops blinking the set
temp. and shows the current
measured temperature.