4.2 Basic calibration with a signal output from 0 ... 10 V
1.

Select the temperature range on the printed circuit board
using the DIP switch, see table on the module or pages 23
and 24.
The DIP switches 1, 2 and 3 set the minimum input tempera-
ture

ϑ

min.

The span (difference between minimum and maximum input
temperature) is set using the DIP switches 4, 5 and 6.

2.

Professionally install module.

3.

Add 1 % of the temperature span to the selected “minimum
temperature and set this value on a PT 100 simulator (when
using a precision resistance decade, where necessary ob-
serve DIN IEC 751 conversion table from °C to

!) and cali-

brate the signal conditioner output signal to 0.100 V using the
null potentiometer. (The potentiometers are located behind
the hinged cover).

4.

Set the “maximum” selected temperature on a PT 100 simu-
lator (or precision resistance decade) and calibrate the signal
output to 10.000 V using the Span potentiometer.

5.

Repeat steps 3 and 4 (approx. 2-3 times), until the required
accuracy is achieved.