Warning – UVP PS-11 Pen-Ray Power Supplies User Manual
Page 2
![background image](/manuals/574838/2/background.png)
Pen-Ray Power Supplies
PS-1, PS-2, PS-3, PS-4, PS-9 and PS-11
Operating Instr
uctions
WARNING
Please read operating instructions before turning unit on.
INTRODUCTION
The Pen-Ray
Power Supplies are made specifically for the Pen-Ray Lamps.
These
power supplies provide proper voltage and current for starting and operat-
ing the lamps. Mating
recommendations are available to provide optimum energy
output from and maximum life span of the Pen-Ray lamps.
DESIGN
Each power
supply is designed to be maintenance free. There is an internal
circuit protection
fuse which is designed to prevent mains circuit shut down in the
unlikely case
of a catastrophic internal failure. The connector must be used with
the mating
connector supplied with the Pen-Ray lamps. The connector is certified
to UL
and CE
compliance for use with this power supply
. The
power supply carries
the CE
, ETL
and ETLc
marks.
OPERATING INSTRUCTIONS
The operation of the PS series power supply is
as follows:
1. Make
sure the switch is in the off position.
2. Connect
the Pen-Ray lamp to the connector making sure the connector is
completely
inserted into its mating part.
3. Plug
the unit into proper mains power supply.
4. Turn
on the unit and the lamp should immediately light up.
5. Always
unplug the unit when disconnecting or connecting the lamp.
MAINTENANCE
Keep the connector clean and free from debris, oil, fluids, etc. If
the cord should
become frayed or cracked
, it
should be replaced immediately.
Call UVP for
replacement service.
If connector
arcing should occur due to dirt in the connector or attempting to
operate a Pen-Ray lamp which operates in excess of 300volts
, do not continue
operation. The
unit should be returned to the factory for repair
. (Cleaning the
carbonized area
between the connector pins may restore function to the power
supply).
W ARNINGS
•
Never handle the connector when the unit is in position as
a hazardous shock
condition will exist.
•
Do not
attempt to operate lamps with operating voltages greater than 300 volts
continuously. S
tarting
voltages
do not apply.
•
If the equipment is used in a manner not specified by UV
P, LLC, the protection
provided by the equipment may be impaired.
Unit may become HOT during continuous use.
If internal
circuit protection fuse blows, the unit should be returned to the factory for
service. The internal protection fuse is designed to trip if the primary coil short
circuits for any reason
, protecting the user and the circuit it is plugged into. The unit
may be cleaned with any mild cleaning agent.
Although the powder coated finish is
extremely durable, avoid
acids and caustic substances
. Do not use solvent cleaners
on plastic end caps.
ENVIRONMENT AL
CONDITIONS
•
For indoor use only
•
Temperature
: 5 to 40
O
C
•
Maximum o
r R
elative H
umidity: 80% for temperatures up to
31
O
C decreasing
linearly to 50% relative humidity at 40
O
C
•
Mains supply voltage fluctuations not to exceed
+10%
of the nominal voltage
•
Installation is category II
•
Pollution degree is 2
SPECIFICATIONS
Primary
Model
Primary
Current
Starting
Max. Lamp
Lamp
No.
Vol
tag
e
AMPS
Volta
ge
V
olta
ge
Current
Dimensions
PS-1
115
Vac
/60Hz
0.40
2300Vac
300V
ac
18mA
6.3x3.7x2.3"
PS-1
230Vac/50Hz
0.21
2300Vac
300V
ac
18mA
(160x94x53mm)
PS-2
115
Vac/60Hz
0.40
2000Vac
300V
ac
6mA
6.3x3.7x2.3"
PS-2
230Vac/50Hz
0.21
2000Vac
300V
ac
6mA
(160x94x53mm)
PS-3
115
Vac/60Hz
0.40
2100Vac
300V
ac
10mA
6.3x3.7x2.3"
PS-3
230Vac/50Hz
0.21
2100Vac
300V
ac
10mA
(160x94x53mm)
PS-4
115
Vac/60Hz
1.00
2800Vac
300V
ac
26mA
4.6 x
5.8 x 3.8”
PS-4
230Vac/50Hz
0.55
2800Vac
300V
ac
26mA
(117x147x97mm)
PS-9
115
Vac/60Hz
0.90
2400Vac
300V
ac
48mA
5.8 x
4.6 x 3.6”
PS-9
230Vac/50Hz
0.50
2400Vac
300V
ac
48mA
(147x117x92mm)
PS-11
115
Vac/60Hz
0.40
2300Vac
300V
ac
18mA
5.2x2.2x3.5"
PS-11
230Vac/50Hz
0.21
2300Vac
300V
ac
18mA
(132x56x89mm)
NOTE: 115V
ETL LISTED US/CANADA; 230V CE COMPLIANCE
81-0154-01 Rev
F
Web Site: www.uvp.com
UVP , L
LC
2066 W
. 1
1th Street,
Upland, C
A 91786
• T
el: (909)
946-3197 or
(800)
452-6788
Ultra-Violet Products L
td
, Unit
1, Trinity Hall Farm Estate, Nuffield Road,
Cambridge, CB4
1TG
U
K •
Tel:
+44
(0)
1223 420022