Replacing the uv germicidal or white light tubes, Replacing the fuses – UVP Ultraviolet PCR User Manual
Page 13

Ultraviolet PCR Cabinets and Workstations
Page 13
UV PCR Workstation
34-0073-01
25W 254nm
UV
2
95-0367-xx
34-0087-01
15w White
1
UV2 PCR
Workstation
34-0007-01
8W 254nm UV
1
95-0439-xx
34-0073-01
25W 254nm
UV
2
34-0087-01
15w White
1
Accessories
Part Number
Table, stainless steel top
98-0077-01
Shelf, long 27”
20-0970-01
UV/Air Circulator Module (for 95-0367-xx)
97-0157-01
UVX Radiometer
97-0015-02
UVX Sensor (UVX-25)
97-0016-01
J-225 Meter
97-0004-01
Spectacles, UV blocking (UVC-303)
98-0002-01
Goggles, UV blocking (UVC-503)
98-0002-02
Face Shield, UV blocking (UVC-803)
98-0002-04
Replacing the UV Germicidal or White Light Tubes
Turn the unit power off and unplug the unit from the outlet. Lift up the panel door and locate the
tubes. The tube has two prongs on each end that first fit into a socket then rotate into place.
Carefully hold the tube on the end and insert the tube into the socket.
Twist the tube 90° until the tube locks into place.
Cabinet:
Outer (White Bulb) – White light (8 watt)
Inner (Clear Bulb) – UV (8 watt)
Workstation:
Outer (Clear Bulb) – UV
Inner (White Bulb) – White light
Replacing the Fuses
Fuses are located on lower left side of the
system. To remove the fuse, use a flathead
screwdriver to turn the fuse counter
clockwise. Pull the fuse out. Insert the new
fuse and using the screwdriver, turn the fuse
clockwise to lock.