Service Procedures
Return Procedure
A
Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product.
Replacing the Battery
The Luminescence Calibration Standard is powered by a standard 9V alkaline battery. To replace the
battery:
1. Locate the Battery Compartment Door on the top of the unit. Remove the Door by removing the screw
holding the door in place (Fig 1).
2. Pull the battery slightly out of the unit (Fig 2). Then, disconnect the Battery Terminal Connector from the
battery (Fig 3).
3. Once the Battery Terminal Connector is fully disconnected, remove the 9V battery from the unit. Then,
slide a new battery in with the positive (smaller) terminal facing toward the front of the unit (Fig 3).
4. Place the Battery Terminal Connector on top of the battery. Press down on the Connector until it snaps
into place on both terminals (Fig 4).
5. Place the Battery Compartment Door back onto the unit, then tighten the screw to hold the door in place
(Fig 1).
Positive Battery
Terminal