beautypg.com

System components and operation, Operating the unit, Usage notes – UVP Luminescence Calibration Standard User Manual

Page 3

background image

LED Calibration Standard

3

System Components and Operation

Refer to the packing slip and the picture below for specific parts and components included with the system.

Operating the Unit

1. Turn the power switch to the ON (

I) position. The Status Indicator Light will glow yellow. The LED Light in the

Aperture may not be visible to the naked eye in strong ambient light or with an emission filter selected.

2. Wait approximately one minute for the Status Indicator Light to extinguish. This indicates that the LED output

has stabilized and that the Standard is ready for use.

3. Rotate the Filter Selection Wheel to the appropriate position for the application. The appropriate position is

the one in which the LED Aperture’s output most closely matches the comparison source (such as a

chemiluminescent sample) without being oversaturated. The relative light output in each position is:

Setting 1: 100%

Setting 2: 10%

Setting 3: 1%

Setting 4: 0.1%

4. Place the Luminescence Calibration Standard in the imaging device along with the comparison source.
5. Adjust the camera settings using UVP’s VisionWorks

®

LS software so that the comparison source and

Focusing Aid are both in focus. Then, capture the image.

6. Use the VisionWorksLS “Area Density” analysis tool to compare the relative intensities of the Standard’s

LED Light Aperture and comparison source.

7. Capture additional images with other comparison sources if desired.
8. Once testing is complete, turn the Standard’s power switch to the OFF (

O) position.

Usage Notes

1. For image intensity comparisons to be meaningful, it is important to keep as many experimental variables as

constant as possible. To do so:

Maintain constant camera exposure times and lens aperture settings;

Keep the camera and imaging environment temperature as constant as possible; and

Place the LED aperture in the same location relative to the camera lens between trials to maintain

a static light angle and path.

2. A second comparison source is not required for all experiments. For example, a comparison of lens

efficiencies is possible by simply using the Luminescence Calibration Standard and switching lenses.

LED Light Aperture

Status Indicator

Light

Neutral Density

Filter Selection

Wheel

Power Switch

Power Indicator

Lights

Focusing Aid

Battery

Compartment Door

This manual is related to the following products: