INTRODUCTION
UVP Crosslinkers offer researchers an instrument to quickly, safely and efficiently expose samples to a
controlled amount of ultraviolet radiation. CL-1000 series features a lay-down type drawer while the CX-
2000 has a pull out drawer. Exposure of samples to UV provides for the following:
• Crosslinking of DNA or RNA to nitrocellulose, nylon or reinforced nitrocellulose.
• PCR sample contamination control.
• Photoniking of DNA
• Testing RecA function
• Rapid site mapping
• UV sterilization and sanitization
• UV curing
The Crosslinkers are designed to measure and control the ultraviolet (UV) radiation within the exposure
chamber. A unique UV sensor continually measures the UV energy and automatically adjusts to
variations in UV intensity that occurs as the UV tubes age. This same UV sensor feedback
measurement system allows you to set UV sample exposure, which automatically deactivates the UV
sources when the set UV energy dose has been achieved.
IMPORTANT SAFETY INFORMATION
Ultraviolet Crosslinkers are a powerful source of ultraviolet radiation. Even though they are not easily
accessible, do not attempt to disengage or override the internal safety interlocks. Exposure to the UV
radiation may result. If the UV sources remain on when the door is open, the unit is malfunctioning and
use should be discontinued until the unit is serviced. Do not expose unprotected eyes or skin to UV
radiation. Always disconnect the UV Crosslinker from its electrical supply before servicing.
DESCRIPTION AND SPECIFICATIONS
The dimensions for:
CL-1000 Series UV Crosslinkers:
External:
8.75"H x 13.75"D x 15.75"W (22.2 x 34.9 x 40.0cm)
Interior chamber dimensions:
5"H x 12"D x 10"W (12.7 x 30.5 x 25.4cm)
Weight:
15.5 lbs. (7.5 kg)
Tubes:
5 x 8 watt UV dual bipin discharge type
CL-1000 Model:
254nm UV
CL-1000L Model:
365nm UV
CL-1000M Model:
302nm UV
CX-2000 Crosslinker
254nm UV
External:
15D x 17.5W x 9.75H in. (381 x 445 x 248 mm)
Internal Chamber:
11.3D x 11.4W x 3.6H in. (287 x 290 x 91 mm)
Features
• Microprocessor controlled/UV sensor feedback system
• Multiple set functions-
preset UV energy exposure
preset UV time exposure
user set UV energy exposure
user set UV time exposure
• Maximum UV energy exposure setting of 999,900 microjoules per cm2
• Maximum UV time exposure setting of 999.9 minutes