Service procedures, Bulb replacement – UVP BioLite Xe User Manual
Page 13
BioLite Xe MultiSpectral Light Source
13
Service Procedures
Bulb Replacement
The BioLite Xe xenon light bulb is a lamp assembly which includes a mounting plate and bracket. The
assembly is installed through the bottom of the unit as shown:
1. Unplug the power cable and USB cable from the back of the BioLite Xe and allow the bulb at least
½ hour to cool.
2. The lamp assembly will have a wire coming from each end of the xenon bulb. One wire has a male
connector on the end and the other wire has a female connector. Attach the male connector from
the lamp to the female connector from the BioLite Xe unit, and vice versa.
CAUTION: DO NOT TOUCH THE GLASS OF THE BULB!
3. Insert the lamp assembly into the bottom of the BioLite Xe, aligning the three all-thread screws on
the bottom of the BioLite Xe with the three holes in the lamp assembly mounting plate.
4. Reconnect the plate to the bottom of the unit using the three supplied nuts.
Mounting Plate
Bracket
Light Bulb
Three
Mounting
Nuts
Attach the Male
and Female
Connectors