SATEC PTS175 Manual User Manual
Page 12
![background image](/manuals/572141/12/background.png)
Appendix A Technical Specifications
PTS175 MV Advanced Analyzer
12
Labels: Polycarbonate (UL94-V0)
Ingress Protection: IP65
A.3
Power Supply inputs
120/230 VAC-DC Input:
Rated input: 85-265VAC 50/60/400 Hz, 88-290VDC, Burden 9VA
Lightening withstanding: 10,000A
12 VDC Input:
Rated input: 12 VDC ± 10%, 10W
Lightening withstanding: 1000A
All Power supply sources must be grounded to earth
A.4
Input Ratings
Voltage Inputs
Primary Measuring range: 11-33 KVAC line-to-line, 6.35-20 KVAC line-to-neutral
Input impedance: 1000 k
Current Inputs
Primary Measuring range: up to 600A
Sampling Rate measurement
128 samples/cycle.
Lightening withstanding: 10,000A
A.5
Real-time Clock
Accuracy: typical error 1 minute per month @ 25
C
Typical clock retention time: 2 years
A.6
Standards Compliance
Accuracy:
Meet IEC62053-22, class 1
Electromagnetic Immunity:
Comply with IEC 61000-6-2:
IEC 61000-4-2 level 3: Electrostatic Discharge
IEC 61000-4-3 level 3: Radiated Electromagnetic RF Fields
IEC 61000-4-4 level 3: Electric Fast Transient
IEC 61000-4-5 level 4: Surge
IEC 61000-4-6 level 3: Conducted Radio Frequency
IEC 61000-4-8: Power Frequency Magnetic Field
Electromagnetic Emission:
Comply with IEC 61000-6-4: Radiated/Conducted class A
Comply with IEC CISPR 22: Radiated/Conducted class A
Safety/Construction:
Meets UL/IEC 61010