beautypg.com

Overview, Specifications – RKI Instruments 65-2499RK User Manual

Page 5

background image

65-2496RK/65-2499RK CO Detector • 1

Overview

This manual describes the 65-2499RK carbon monoxide (CO) detector. This manual also
describes how to install, start up, maintain, and calibrate the CO detector when used with
a gas monitoring controller. A parts list at the end of this manual lists replacement parts
and accessories for the CO detector.

The 65-2499RK CO detector includes the 65-2496K CO detector and a junction box. This
manual may also be used for the 65-2496RK CO detector which does not include a
junction box and is normally mounted in one of a controller’s conduit hubs. If you are
using a 65-2496RK CO detector, disregard all references to the junction box and junction
box terminal strip.

Specifications

WARNING: Do not use this product in a manner not specified in this instruction

manual.

Table 1 lists specifications for the 65-2496RK and 65-2499RK CO detectors.

WARNING: When using the 65-2496RK/65-2499RK, you must follow the instructions

and warnings in this manual to assure proper and safe operation of the
65-2496RK/65-2499RK and to minimize the risk of personal injury. Be sure
to maintain and periodically calibrate the 65-2496RK/65-2499RK as
described in this manual.

Table 1: Specifications

Target Gas

Carbon Monoxide (CO)

Sampling Method

Diffusion

Detection Range

0 to 300 PPM (parts per million)

Response Time

90% in 30 seconds

Accuracy

± 5% of reading or ± 5 ppm CO (whichever is greater)

Operating Temperature

-4°F to 113°F (-20°C to 45°C)