RKI Instruments 65-2432RK-05 User Manual
Page 14

10 • 65-2432RK-05 CO Transmitter
•
Calibrate the transmitter.
•
If the fail condition continues, replace the CO sensor.
•
If the fail condition continues, contact RKI for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
•
The transmitter responds slowly or does not respond during the monthly response
test.
•
Unable to accurately set the zero or response reading during the calibration
procedure.
•
The transmitter requires frequent calibration.
NOTE:
Under “normal” circumstances, the transmitter requires calibration once every
three months. Some applications may require a more frequent calibration
schedule.
Probable causes
•
The calibration cylinder is low, out-dated, or defective.
•
The transmitter is malfunctioning.
Recommended action
1.
Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2.
If the calibration/response difficulties continue, replace the CO sensor as described
later in this section.
3.
If the calibration/response difficulties continue, contact RKI Instruments, Inc., for
further instruction.
Replacing Components of the CO Transmitter
This section includes procedure to replace the CO sensor and amplifier. A procedure to
replace the entire detector assembly is at the end of this section. In most cases, it is not
necessary to replace the entire detector assembly.
Replacing the Sensor
CAUTION: The sensor contains electrolyte which is a dilute acid. Do not disassemble the sensor
when replacing it with a new one. If sensor electrolyte comes in contact with your
skin, wash affected area thoroughly with soap and water.
1.
Turn off the controller.
2.
Turn off or unplug incoming power to the controller.
3.
Unscrew the bottom section of the CO detector housing from the top section.
4.
Unplug and remove the CO sensor with the boot and charcoal filter attached.
5.
Remove the rubber boot and charcoal filter from old sensor.
6.
Install the rubber boot with charcoal filter onto the replacement sensor’s face.
7.
Carefully plug the replacement sensor into the socket pattern that is located in the top
section of the detector housing.