beautypg.com

RKI Instruments 65-2432RK-05 User Manual

Page 13

background image

65-2432RK-05 CO Transmitter • 9

test point labeled TP-.

8.

Use the following formula to determine the correct test points output for the test
sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a test sample of 50 PPM CO and a fullscale setting of
300 PPM, the correct output is 167 mV.

167 (mV) = (50/300) X 400 +100

Performing the response test

1.

Turn the regulator on/off knob counterclockwise to open it. The sample will begin to
flow.

2.

Allow the gas to flow for two minutes, then verify that the reading is within ± 20% of
the response reading you determined earlier.

NOTE:

If the readings are not within ± 20% of the correct response reading, calibrate the
affected transmitter(s) as described in the Calibration section of this manual.

3.

Turn the regulator on/off knob clockwise to close it.

4.

Unscrew the regulator from the calibration cylinder.

5.

Unscrew the calibration cup from the CO detector.

6.

Remove the voltmeter leads from the amplifier test points.

7.

Reinstall the junction box cover.

8.

When the controller display reading falls below the alarm setpoints, return the
controller to normal operation.

Quarterly

Calibrate the CO transmitter as described in the Calibration section of this manual.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO transmitter.

NOTE:

This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the
controller.

Fail Condition

Symptoms

The controller indicates a fail condition.

Probable causes

The transmitter wiring is disconnected or misconnected.

The transmitter’s zero reading is low enough to cause a fail condition.

The transmitter is malfunctioning.

Recommended action

Verify that the transmitter wiring is correct and secure.