beautypg.com

RKI Instruments 65-2424RK User Manual

Page 14

background image

10 • 65-2424RK H

2

S Transmitter

7.

Secure the cover to the enclosure.

8.

Turn on or plug in incoming power at the power source end.

9.

Turn on the controller.

CAUTION:

Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.

10. Calibrate the replacement sensor as described in the Calibration section of this

manual.

Replacing the Amplifier PCB

1.

Turn off the controller.

2.

Turn off or unplug incoming power at the power source end.

3.

Remove the enclosure’s cover by unscrewing the four cover screws, one at each
corner.

4.

Disconnect the wires from the interconnect terminal strip.

5.

Unplug the H

2

S sensor from the amplifier PCB to access one of the PCB mounting

screws.

6.

Unscrew and remove the two screws that secure the amplifier PCB to the enclosure.
The screws are near the top and bottom of the amplifier PCB.

7.

Remove the amplifier PCB.

8.

Place the new amplifier PCB in the same position as the one you removed in step 6.

9.

Use the two screws you removed in step 5 to secure the PCB to the enclosure.

10. Reconnect the wiring from the controller to the interconnect terminal strip as shown

in Table 2 and Figure 3.

11. Carefully plug the H

2

S sensor with rubber boot and charcoal filter into the socket

pattern on the amplifier PCB.

NOTE:

Match the sensor’s male pins with the four female sockets as you plug the sensor
into the sockets.

12. Secure the cover to the enclosure.

13. Turn on or plug in incoming power at the power source end.

14. Turn on the controller.

CAUTION:

Allow the sensor to warm up for 5 minutes before you continue with the next step.

Table 2: Reconnecting the

H

2

S Transmitter to a Controller

Amplifier PCB

Interconnect Terminal Strip

Controller

Transmitter Terminal Strip

(typical)

FB

4 - 20 (FB)

+V

+ V (10 VDC - 30 VDC)