beautypg.com

RKI Instruments 65-2424RK User Manual

Page 13

background image

65-2424RK H

2

S Transmitter • 9

Recommended action

Verify that the transmitter wiring is correct and secure.

Calibrate the transmitter.

If the fail condition continues, replace the H

2

S sensor.

If the fail condition continues, contact RKI for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms

The transmitter responds slowly or does not respond during the monthly response
test.

Unable to accurately set the zero or response reading during the calibration
procedure.

The transmitter requires frequent calibration.

NOTE:

Under “normal” circumstances, the transmitter requires calibration once every
three months. Some applications may require a more frequent calibration
schedule.

Probable causes

The calibration cylinder is low, out-dated, or defective.

The transmitter is malfunctioning.

Recommended action

1.

Verify that the calibration cylinder contains an adequate supply of a fresh test sample.

2.

If the calibration/response difficulties continue, replace the H

2

S sensor as described

later in this section.

3.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for
further instruction.

Replacing Components of the H

2

S Transmitter

This section includes procedures to replace the H

2

S sensor, amplifier PCB, and charcoal

filter.

Replacing the Sensor

1.

Turn off the controller.

2.

Turn off or unplug incoming power at the power source end.

3.

Remove the enclosure’s cover by unscrewing the four cover screws, one at each
corner.

4.

Unplug the H

2

S sensor from the amplifier PCB.

5.

Remove the rubber boot and sealing spacer from the old sensor and install onto the
replacement sensor.

6.

Carefully plug the replacement sensor into the socket pattern on the amplifier PCB.

NOTE:

Match the sensor’s male pins with the four female sockets as you plug the sensor
into the sockets.