Calibration – RKI Instruments 65-2335RK User Manual
Page 18

14 • 65-2335RK CO Transmitter
then more frequent calibration than every 3 months may be necessary.
Calibration
This section describes how to calibrate the CO transmitter. It includes procedures to
prepare for calibration, set the zero reading, set the response reading, and return to
normal operation. It describes the test using a calibration kit that includes a calibration
cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI
Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
Preparing for Calibration
NOTE:
Calibrating the CO transmitter may cause alarms. Be sure to put the controller
into its calibration program or disable external alarms before calibrating.
1.
Unscrew and remove the junction box cover.
2.
Set a voltmeter to measure in the millivolt (mV) range.
3.
Plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the red (+) test point; plug the negative lead into the black
(-) test point.
4.
Use the following formula to determine the correct test points output for the
calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50 PPM CO and a fullscale setting of
300 PPM, the correct output is 167 mV.
167 (mV) = (50/300) X 400 +100
5.
Screw the calibration cup onto the detector housing.
Setting the Zero Reading
NOTE
If you can verify that the CO transmitter is in a fresh air environment, you do not
need to apply zero air to the detector before adjusting the zero reading.
1.
Screw the regulator into the zero air calibration cylinder.
2.
Use the sample tubing to connect the regulator to the calibration cup.
3.
Turn the regulator knob counterclockwise to open the regulator.
4.
Allow the gas to flow for two minutes, then verify a reading of 100 mV (± 2mV). If
necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (± 2mV).
5.
Turn the regulator knob clockwise to close the regulator.
6.
Unscrew the regulator from the zero air calibration cylinder.
7.
Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
1.
Screw the regulator into the calibration cylinder. Verify that the calibration gas is
representative of the transmitter’s target gas.
2.
Turn the regulator knob counterclockwise to open the regulator.