Appendix D: Carbon Dioxide Sensors • 68
9. With the Single Calibration menu displayed, press the SHIFT/▼
button until the prompt is next to the last channel, then press the
SHIFT/▼ button again. The
ESCAPE
message displays.
10. Press the POWER/ENTER button to return to the Calibration
menu.
11. Press the SHIFT/▼ button to place the prompt next to the
NORMAL OPERATION
menu option, then press the
POWER/ENTER button to return to the normal screen.
Replacing Carbon Dioxide Sensors
Return the Eagle to RKI Instruments, Inc. for replacement of the
carbon dioxide sensor when:
• The carbon dioxide channel cannot be calibrated correctly.
• The carbon dioxide reading cannot be set to zero by the Demand
Zero command or zero potentiometer.
Parts List
Table 12 lists part numbers for replacement parts and accessories of
the Eagle’s carbon dioxide sensors.
Table 12: Parts List: Carbon Dioxide Sensors
Part Number
Description
33-6010RK-01
Scrubber, carbon dioxide
81-0070RK-03
Calibration cylinder, 103-liter (2000 ppm CO
2
)
81-0071RK-03
Calibration cylinder, 103-liter (5000 ppm CO
2
)
81-0072RK-03
Calibration cylinder, 103-liter (2.5% CO
2
)
81-0073RK-03
Calibration cylinder, 103-liter (15% CO
2
)
81-1054RK
Regulator, demand-flow type (for 58- and 103-liter calibration
cylinders)