RKI Instruments GasWatch 2 Manual User Manual
Page 28

23 • Calibration
GasWatch 2 Operator’s Manual
8. Let the gas flow for one minute and then press the POWER/MODE
button. The unit will adjust the span (GW-2C or GW-2H) or zero (GW-
2X) based on the calibration value that was saved in step 5 above.
9. Turn the regulator knob clockwise to close the regulator.
10. Remove the regulator from the gas cylinder and the calibration adapter
from the unit. As soon as the unit makes the calibration adjustment, it
will begin its startup sequence and then enter Measuring Mode.
NOTE: If the gas reading is high enough (low enough for the GW-2X)
when the unit enters Measuring Mode, an alarm condition will
occur. Reset the alarm using the POWER/MODE button when the
gas reading falls below (rises above for the GW-2X) the alarm
point.
- Eagle Series (89 pages)
- Eagle Series (66 pages)
- 01 Series Manual (33 pages)
- 01 Series Quik Start (2 pages)
- OX-94 (13 pages)
- FP-3 (16 pages)
- GasWatch 1 (18 pages)
- GasWatch 2 Quik Start (2 pages)
- GP-01 (34 pages)
- GP-01 with self resetting alarms (33 pages)
- GP-204 (7 pages)
- NP-204 (13 pages)
- RI-215A (24 pages)
- SC-01 (68 pages)
- SC-01 (45 pages)
- Data Cal 2000 (88 pages)
- EAGLE 2 Manual (309 pages)
- Eagle 2 Quik Start (2 pages)
- GX-94 (3 pages)
- GX-94 (52 pages)
- SDM-E2 (224 pages)
- SDM-E2 (89 pages)
- Eagle 1 Manual (89 pages)
- Eagle 1 Quik Start (2 pages)
- Inert Gas Testing Eagle with Two Pumps (66 pages)
- Gas Tracer (234 pages)
- Gas Tracer 10,000 ppm (120 pages)
- GX-2001 Manual (45 pages)
- GX-2001 Quik Start (2 pages)
- GX-2001 Canadian Version (43 pages)
- GX-2003 Manual (94 pages)
- GX-2003 Quik Start (2 pages)
- GX-2009 Manual (73 pages)
- GX-2009 Quik Start (2 pages)
- GX-2009 Supplement (2 pages)
- GX-2012 Manual (185 pages)
- GX-82 (22 pages)
- GX-82A (43 pages)
- GX-86 (22 pages)
- GX-86A (39 pages)
- GX-8000(PT0-098) (52 pages)
- GX-8000(H4-0050) (23 pages)
- RX-516 (36 pages)
- GX-2009 (1 page)