General
Cleaning
For thorough, gentle and time-saving cleaning of your ana-
lytical sieves, we recommend the Retsch ultrasonic baths.
Ask for our special publication "Care and cleaning of analytical
sieves" free-of-charge.
Do not clean the AS300 control with running water.
Danger to life from electrical shock
Only use a rag moistened with water.
Never use solvents.
Maintenance
If you use your AS300 control for quality control purposes, it
should then be calibrated regularly in accordance with DIN EN
ISO 9000 ff. In this context, please contact or your dealer or
speak directly to Retsch GmbH.
Otherwise the AS300 control is virtually maintenance-free.
Replacing fuses
B
Two glass fuses SPT 6.3A (5x20mm) are required.
Unplug the power at the socket
Pull out fuse holder B
Exchange the fuse
Re-insert fuse holder B
Accessories
Sieve clamping unit standard for analytical sieves
305mm (12“)
Standard sieve set 305mm (12“) 50mm high,
consisting of 7 analytical sieves to DIN 3310/1
0.63 + 1.25 + 2.5 + 5 + 10 + 20 and 31.5mm
plus base pan
Standard sieve set 305mm (12“) 50mm high,
consisting of 7 analytical sieves to ASTM
30 + 16 + 8 + 4 mesh + 3/8“ + 3/4“ and 1 1/4“
plus base pan