Intermittent – continuous operation
7
8
9
10
11
When the AS300 control is switched on, intermittent operation is ini-
tially disabled.
Enabling intermittent mode of 10-99 sec. :
Press key 10
The LED above key 10 illuminates
Display 8 shows 10 sec.
Pressing key 9 raises the interval period up to 99 sec.
If 99 sec is overshot, start again with 10 sec.
Key 7 reduces the interval period down to 10 sec.
If 10 sec. is undershot, start again with 99 sec.
Disabling intermittent mode :
Press key 11
The LED above key 11 illuminates
Display 8 is extinguished
Memory – saving and calling-up sieve parameters
2
5
8
10 11
13
The memory function, thus saving or calling-up preselected sieve pa-
rameters, is only available in standby mode. You are able to change
and overwrite program places P1 to P9 at any time.
After the AS300 control has been turned on, display 2 shows "on".
The next program place P1 to P9 is reached by pressing the PROG
key. on reappears in the display after program place P9.
on = the sieve parameters 5/8/10-11 and 13 can be freely
set here.
or
P1 to P9 = sieve parameters can be stored and called-up here
Start = sieving process is started with the sieve parameters
stored in program place P1 to P9.
Occupying storage space
Press the PROG key until the required storage place P1-P9
has been reached.
Press the SET key, all displays flash
Set sieve parameter 5/8/10-11/13
The programming mode can be aborted by pressing the PROG button
– the values are not stored.
Press SET button – the values are stored
Flashing of the display stops, the adjusting interlock is activated, the
sieving parameters are stored.
If the AS300 control is actuated with a PC and the "EasySieve
®
"
sieve software, display 2 shows "ES". No manual changes to the sieve