10
Operation
Operating elements and their use
A Cold device plug combination
C Fuse drawer with glass fuses
B Serial connection
D Protective cap for the serial connection
E ON / OFF switch, left bottom
1 2
3
4 5 6
7
8
9
10 11 12 13 14 15 16
1
PROG – button for calling preset sieving parameters P1-P9 or on and for ending pro-
gramming without storing the parameters.
2
Display memory shows the selected programme P1-P9 or on. If the AS 200 is driven by
the EasySieve
®
softwear programme ES appears in the display.
3
SET – button for activating the setting mode for the preselected programme positions P1-
P9 and for storing the parameters after programming.
4
▬ button reduces the amplitude 0.20 – 2.20mm, precision ±0.1mm, or sieve bottom
acceleration in „g“
5
Display amplitude shows the amplitude (2 x amplitude) of 0.20 – 2.20mm or sieve bot-
tom acceleration in „g“.
6
+
button increases the amplitude, 0.20 – 2.20mm, precision ±0.1mm, or sieve bottom
acceleration in „g“.
7
▬ button reduces the interval time, 10 – 99sec.
8
Display interval shows the preselected interval time, 10 – 99sec.
9
+
button increases the interval time, 10 – 99sec.
10
Button switches interval operation ON, left LED lights up.
11
Button switchces interval operation OFF, right LED lights up.
12
▬ button reduces the sieving time, 1 – 99min.
13
Display time shows the preselected sieving time, 1 – 99min.
14
+
button increases the sieving time, 1 – 99min.
15
START button starts the sieving process and the green LED lights up.
16
STOP button stops and ends the sieving process and the red LED lights up.