Safety regulations for the as200, Extract from the individual chapters – Retsch AS 200 User Manual
Page 30
31.07.2012
Retsch GmbH
30
Safety regulations for the AS200
Extract from the individual chapters
Subject
Action
Hazards
Vibration height
Vibration height >2.5 mm in the
AS200 basic and digit models
Armature strikes the magnets;
mechanical components could
be damaged.
Shipping
During shipment the AS200 is not
to be subjected to impact or
vibration; do not throw the
package.
Electronic
and
mechanical
components could be damaged.
Temperature
fluctuation
In
case
of
temperature
fluctuations, protect the AS200
against condensation.
Electronic components could be
damaged.
Scope of supply
If the shipment is not complete or
is damaged, you must notify the
forwarder and Retsch GmbH
immediately (within 24 hours).
Under certain circumstances
compensation cannot be paid
for claims which are lodged
later.
Ambient
temperature
Temperature below 5°C
Temperature above 40°C
Electronic components could be
damaged.
Humidity
Exceeding
80%
R.H.
at
temperatures up to 31°C
Electronic components could be
damaged.
Installation
Shipping nuts not removed.
Mechanical
and
electronic
components could be damaged.
Electrical
connection
Local power supply does not
correspond to the values given on
the data plate.
Mechanical
and
electrical
components could be damaged.
Sieving aids
Stretching the sieve weave due to
excessive charge weight.
The precision of the analytical
sieve can be impaired.
Sieved product
distribution
Use the long threaded rods when
working with only one to three
sieves.
Overly long rods, extending
above the stack, will interfere
with distribution of the product
due to their intrinsic resonance.
Wet sieving
Never place the AS200 directly in
a basin for operation
Fatal electrical shock hazard.
Line power socket not protected
with a ground fault interrupter.
Fatal electrical shock hazard.
Water volume too great.
Water backing up in the sieve
stack can cause overloading
and thus may damage or
destroy the sieve weave.
Cleaning
Never clean the AS200 with
running water.
Fatal electrical shock hazard.