Operation for the AS200 basic
Switching on and off
The main switch J is located at the left-front on the AS200; see
Fig. 14.
*
Turn the main switch on.
*
The ON LED lights, Fig. 15.
The AS200 is now ready for operation.
Fig. 14
Start - Stop
Fig. 15
Start: Fig. 15
*
Set knob UV2 for continuous operation
.
Stop (standby function): Fig. 15
*
Set knob UV2 to I.
Vibration height
If you desire to work with vibration heights above 2.5
mm, we recommend running the AS200 up to this value
slowly. Depending on the loading situation, resonance
(harmonic reinforcement) effects can be induced,
causing the armature and the yoke of the magnet to
make contact.
Setting the vibration height (twice the
amplitude), from 0 to 3.00 mm: Fig. 15
*
Turn the MN2 to the left to reduce the vibration height, to
the right to increase.
The values shown on the potentiometer scale are for
orientation purposes only. There is no absolute
relationship to the actual vibration height and thus this
scale cannot be used as an aid in reproducing previous
values.