Retsch BB 100 User Manual
Page 18
2/7/2012
Retsch GmbH
18
Replacing the breaker jaws
Fig. 10
The stationary breaker arm S can be removed to provide
sufficient space for dismantling the machine.
Fig. 10
Unscrew the machine screws E.
Swing back the fill hopper B.
Grasp the breaker arm S at the handle R and pull out the
pin Q.
Lift out the breaker arm S.
Unscrew the machine screws H.
Remove the clamping shims G.
Replace the breaker jaws J.
Assemble in reverse order.
Set the gap width as described in this instruction manual.
!
The fill hopper B may be swung back (by unscrewing the two
machine screws E) only by trained personnel and only for
cleaning purposes. Operating the BB 100 without these
machine screws E in place and tightened down properly is
absolutely prohibited.
Hazard of crushing fingers and hands!!
Copyright
Reproducing or distributing this manual, utilizing or distributing
the contents is permitted only with the express consent of
Retsch GmbH.
Non-compliance will subject violators to claims for
damages.
Modifications
Subject to modification without prior notice.