Page 22, All thermocouple connections – Practical Instrument Electronics 820-ELITE User Manual
Page 24

Page 22
Thermocouple Source
Choose this function to provide a simulated thermocouple signal
into controllers, temperature transmitters, indicators or any input
devices that measure thermocouple sensors.
Move the power switch
w
to SOURCE then Double Click the
EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select T/C for the FUNCTION, °F or °C for the UNITS,
T/C Type (J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) or
P (Platinel II) and internal COLD JUNC ON or OFF (ON is the
default).
Connect the PIE 820-ELITE to the inputs of the device being
calibrated using the proper type of thermocouple wire via the
miniature thermocouple socket.
Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob
e
. Press and turn the knob for
faster dialing with 10.0° increments.
All thermocouple connections
SOURCE
HI
SET
LO
OFF
MAX
READ
MIN
READ
OVERLOAD
Hz SYNC
Double Click
Menu
Push & Turn
for
Fast Dialing
Push & Hold
to
Store/Step
READ
TYPE K °C
820.0
LO
mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure
Instrument with T/C Input
Controller
Temperature Transmitter
Temperature Indicator
Temperature Trip or Alarm
Thermocouple Wire