Typical error conditions, Page 17, Ma pwr meas – Practical Instrument Electronics 820-ELITE User Manual
Page 19: Using ground leak detection, Process indicator

Page 17
The PIE 820-ELITE is
supplying the loop voltage.
There is a control loop
error. This may be a
transmitter (set for upscale
burnout) with a bad or
missing sensor, or a short
in the loop.
The PIE 820-ELITE is
supplying the loop voltage.
A calibrated transmitter is
limiting the loop current to
12.00 mA. An additional
0.28 mA is not controlled
by the transmitter and is
leaking somewhere in the
loop.
LEAK ERROR =
OVER RANGE
PWR MEAS
LEAK ERROR = 00.28
12.280
mA
PWR MEAS
Typical Error Conditions