Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 7

mA OUT, % OUT (Percent of 4 to 20 mA)

Choose this function to simultaneously supply power to a 2 Wire Transmitter while displaying the 4
to 20 mA output of the transmitter.

1) Disconnect one or both input wires from

the device to be calibrated.

2) Select

“mA” or “% 4 to 20mA” with slide

switch

q

.

3) Select

“SOURCE” using slide switch

w

.

4) Press and turn the knob

r

clockwise

several times until full scale output (24.000
mA/125.00%) is obtained. The display will
indicate

“PWR MEASURE”.

5) Connect the red source lead of the PIECAL

434 to the plus (+) input of the device and
the black source lead to the minus (-).

The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter. The current
passed by the transmitter will be accurately displayed by the PIECAL 434. Calibrate the transmitter
in the usual manner and disconnect the PIECAL 434. If Ground Leak Detection is enabled the
amount of milliamps leaking in the transmitter or loop is indicated on the display.

Power & Measure 2-Wire Transmitters

Typical

2-Wire

Transmitter

+ IN - REF +OUT-

Transmitter Input

Sensor

Process Signal

Simulated Input

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

OUT

%

25.42