Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 2

q

POWER SWITCH

Select

“mA” to display and calibrate in milliamps.

Select

“% 4 to 20 mA” to display and calibrate

in percent. Select

“READ VDC” to read volts

DC. Return the slide switch to the

“OFF”

position when not in use.

Note:
Percent mode can also be used with chart
recorders, valves or current trips that display in
percent.

100.00% = 20.000 mA

75.00% = 16.000 mA
50.00% = 12.000 mA

25.00% = 8.000 mA

0.00% = 4.000 mA

To convert from Milliamps to Percent:

Percent = (Milliamps - 4) / 0.16

To convert from Percent to Milliamps:

Milliamps = Percent / 6.25 + 4

w

SOURCE / READ / 2 WIRE SWITCH

Select

“SOURCE” to output in mA or percent.

Select

“READ” to read in mA or percent.

Select

“2 WIRE” to simulate a 2 Wire Transmitter.

e

EZ-CHECK™ SWITCH

Instantly output 4.000 mA or 20.000 mA by moving
the EZ-CHECK™ switch to the

“4.00mA” /“0.0%”

position or

“20.00mA”/“100.0%” position. For fast

three point checks select the

“DIAL” position.

The PIECAL 434 will remember the last

“DIAL”

value, even with the power off.

Note: The same

“DIAL” value is stored for both

mA and %. The recalled value will be displayed
in the units selected.

r

DIAL KNOB

Turn the knob to adjust output level. Turn
clockwise to increase the output, counter
clockwise to decrease the output 0.001 mA
(0.01%) at a time. Push and turn the knob for
faster dialing adjusting 0.100 mA (1.00%) at a
time. Press and Hold the knob to start automatic
stepping or ramping. Double click the knob for
stepping and Ground Leak Detection setup.

t

EXTERNAL POWER JACK (Not Shown)

When used with the optional AC Adaptor, the
external power jack will eliminate the drain on
your batteries. This is handy for jobs that require
extended bench use of the PIECAL 434.

See accessories (page 10) for ordering information.

Note: This feature does not charge the batteries,
it only supplies power to the PIECAL 434.

CHANGING BATTERIES
Low battery is indicated by

“BAT” on the display.

Approximately one to four hours of typical
operation remain before the 434 will automatically
turn off. To change the batteries; remove the
rubber boot, remove the battery door from the
back of the unit by sliding the door downward.
This allows access to the battery compartment.
Replace with four (4) “AA” 1.5V batteries being
careful to check the polarity. Replace the battery
door & replace the boot. All stored configuration
options are reset to factory settings when the
batteries are removed.

Note: Alkaline batteries are supplied and
recommended for maximum battery life &
performance.

Basic Operation

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

OUT

mA

20.000

3

5