Press an Up or Down arrow button
or
to change a displayed setting.
The setup menu is as follows:
Indicates that the display alternates between 2 displays.
Temperature Clamp Set Point
When active, the heater will heat the pads until the set point is
reached on the primary pad. Use the Fiine Tune settings to
control the regulation of the temperature clamping.
Alarm Set Point will be 20 degrees C (36 F) above the Set Point.
Fine Tune Temperature Set Point
Fine Tunes thermo-kinetic settings to improve regulation of
temperature Set Point clamping.
Press
to display the Fine Tune settings sub-menu.
Slow Down Temperature Delta
Degrees below Set Point to begin heating slow down.
Heating duty cycle will be reduced based on degrees below
temperature set point and rate of temperature change.
Set to 0 to disable. Default = 10.
Set Point heat clamping duty cycle
base percentage (Pc).
Hold duty cycle percentage is the base heat duty cycle percentage
to trickle charge the heaters to clamp the temperature setting.
Actual percentage will be dynamically adjusted up or down
according to temperature feedback
Set to 0 to disable. Default = 10.
This setting will not appear in Active Mode. C is default setting
Set temperature units: C or F (Celsius or Farenheit). When the
units are changed, all settings will change units without being
converted.
n
Enable Lock Out Mode. Set to 1 to enable. 0 is default setting.
Press
when enabled to set Lock Out code.
nnnn
Lock Out code: use arrow keys to set blinking digit.
Press
or
to move to next digit.
n
Power failure mode. Set to 1 to enable. 0 is default setting.
When Power Failure Mode is enabled, if the heater loses power
while in Active Mode, the heater will power up in Active Mode
when power is restored. Alarm mode will override Active Mode.
nn
RS-232 communications network address. 0 is default setting.
Sets the RS-232 network address for use with a network of
heaters and pumps connected to a computer. 00 is the default
address.