New Era CBL-DUAL-3 RS-232 Dual pumps communications cable (3) User Manual
Part # description contents, Dual pumps cable attachment instructions, Detailed setup instructions

1
D
UAL
P
UMPS
C
OMMUNICATIONS
C
ABLE
Part #
Description
Contents
CBL-DUAL-3 RS-232 Dual pumps communications cable (3’).
RJ-11 Connector
Synchronizes the operation of two pumps in one
of the special communications modes.
Figure 1: Dual Pumps Cable CBL-DUAL-3
Dual Pumps Cable Attachment Instructions
NE-1000 / NE-1000X Series of Syringe Pumps
1)
Turn off power to both pumps
2)
Attach one end of the “Dual Pumps Cable” to the
“To Computer” connector on the back of the
Master Pump.
3)
Attach the other end of the “Dual
Pumps Cable” to the “To Computer”
connector on the back of the
Secondary Pump.
Master Pump (connect to "To Computer")
Secondary Pump (connect to "To Computer")
Detailed Setup Instructions
To select the Master pump control mode, enter Setup on the Master pump by pressing and
holding the "Setup" key (“Diameter” key), until setup is entered. The pump will scroll through
the setup parameters. When the pump address [Ad:00], or other current mode is displayed, press
the left-most up arrow key to scroll through the communications modes:
Mode
Display in
Setup
Description
Address mode
Addr
Communications with PC using cable CBL-PC-PUMP-7.
CBL-DUAL –3 Cable Special Communications Modes
Mode
Display in
Setup
‘X Firmware
Version Only
Description
Dual Pumps
dUAL
Secondary pump duplicates the master pump
Reciprocating Pumps /
Continuous Infusion
rECP
Continuous infusion / auto-refill
Alternating Pump Control
ALtr
Second pump starts when first pump stops.
Next Generation
Continuous Infusion
COnt
Continuous infusion while minimizing flow rate
pauses and drop outs when changing directions
Constant Flow Rate
Gradient
GrAd
Dual pumps inverse linear functions combine for a
constant flow rate
All modes (except Address) are set to 19,200 baud rate. In Address mode, the next setting is the
baud rate setting.
Except for Alternating Pumps mode, the secondary pump should remain in Address mode
[Ad:00] and 19,200 baud rate [1920] setting, which are the factory defaults. With Alternating
Pumps mode, the secondary pump can also be set to Alternating Pumps mode.