1.4
Packing List
Included with the Syringe Heater are the following items:
•
Primary heating pad with 2 releasable plastic ties.
•
Basic Operating Manual.
2.
Overview
The syringe heater controls the power to a heating pad with a temperature sensor. It is controlled from a
microcontroller based system which monitors and regulates the temperature on the heating pad.
Features:
♦
User adjustable Thermo-Kinetic Heat
Clamping for precision regulation of
Set Point temperature.
♦
Maximum temperature of 185 C
♦
Non-volatile memory of all operating
parameters.
♦
RS-232 bi-directional control from a computer with built-
in RS-232 network driver. RS-232 network supports up
to 100 heaters, pumps and other devices from a single
computer port.
♦
Power Failure Mode: Restarts the heater Active Mode
after a power interruption.
♦
Selectable Temperature units of C or F (Celsius or
Fahrenheit)
3.
Setup
♦
Attach the primary heating pad to the larger connector (8 pin) on the end of the heater control box.
♦
Wrap the primary heating pad around the syringe body. Temperature will be measured near the point
where the wires are attached to the heating pad. Secure the heating pad to the syringe with the 2
releasable plastic ties.
♦
Plug in the heater
♦
Use the Up and Down arrow buttons to adjust the temperature Set Point
♦
To start heating the pad, simultaneously press the 2 ACTIVE buttons, SETUP and ENTER/STOP. The
ACTIVE LED will be lit. The heater will begin to regulate the temperature on the pad.
♦
Press the ENTER/STOP button to stop heat regulation and exit Active Mode.
♦
The temperature Set Point can be adjusted while in Active Mode.
Primary Heating
Pad Connector
Secondary
Heating Pad
Connector
Temperature
Sensor
Primary
Heating Pad