3 power failure – Memmert IPS Peltier-cooled incubators User Manual
Page 29

D24034 | Date 10/2014
29
Malfunctions, warning and error messages
Error description
Cause of errors
Rectifying errors
See
Error message E-3 in
the temperature display
TEMP
Set 45.0
°C
E-3
°C
Temperature sensor defec-
tive.
►
Switch off appli-
ance.
►
Remove the cham-
ber load
►
Contact customer
service
page 2
When switching on
the appliance, the start
animation is displayed
in another colour than
white
►
Cyan
:
Not enough storage
space on the SD card
►
Red
:
The system files could
not be loaded
►
Orange
: The
fonts and images could
not be loaded
Contact customer
service
Contact customer
service
Download the firmware
update at memmert.
com and install it
page 2
page 2
6.3 Power failure
In case of a power failure, the appliance operates as follows:
In normal operation
In normal operation
After power supply has been restored, operation is continued with the parameters set. The
time and duration of the power failure are documented in the log memory.
In timer mode
In case of an interruption of the power supply of less than 60 minutes, the current timer is
continued from the point at which it was interrupted. For longer interruptions of the power
supply, all appliance functions are switched off.
In remote control mode:
The previous values are restored.