Construction and description

2.2 Description

The appliance can heat the working chamber up to 70 °C (IPP) or 45 °C (IPS) and cool it down

to 0 ºC (IPP) or 14 °C (IPS). Low-noise, long-life and energy-saving Peltier cooling and heating

technology is used for this. In heating operation, a part of the required energy is extracted

from the surroundings (heat pump principle). Condensation formation during the cooling

down process takes place outside the working chamber on the Peltier element
Optionally, the appliance can be equipped with a light module.

2.3 Material

For the outer housing, MEMMERT deploys stainless steel (Mat.No. 1.4016 – ASTM 430) and

for the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through

its high stability, optimal hygienic properties and corrosion resistance to many (but not all!)

chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with

the materials mentioned. A material resistance table can be requested from the manufacturer.

2.4 Electrical equipment

Operating voltage and current consumption: See nameplate

Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010

Protection type IP 20 acc. to EN 60 529

Interference suppression acc. to EN 55011 class B

Appliance fuse: Safety fuse 250 V/15 A, quick-blow

The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V)

2.5 Connections and interfaces

2.5.1 Electrical connection

This appliance is intended for operation on an electrical power system with a system imped-

ance Z

max

of a maximum of 0.292 ohm at the point of transfer (service line). The operator

max

max

must ensure that the appliance is operated only on an electrical power system that meets

these requirements. If necessary, you can ask your local energy supply company what the

system impedance is.
Observe the country-specific regulations when connecting (e.g. in Germany

DIN VDE 0100 with residual current circuit breaker).