beautypg.com

Memmert HPP Life Constant climate chambers User Manual

Page 14

background image

14

D32802 | Date 10/2014

Operation and control

1

3

4

5

6

7

2

JUMP TARGET

FA

N

Day

FA

N

Evening

Night

Morning

Humidity

Fig. 10

Simulates the day/night cycle in the AtmoCONTROL programme software (example)

1

Jump target for the daily repetition of the programme when reaching L

OOP (7)

2

55 % relative humidity during the entire time (∞)

Jump target for the daily repetition of the programme when reaching L

OOP (7)

Jump target for the daily repetition of the programme when reaching L

OOP (7)

3

Day simulation: 60 % fan speed, 60 % LED lighting and 22°C for 12 hours

55 % relative humidity during the entire time (∞)55 % relative humidity during the entire time (∞)

4

Evening simulation: 20 % fan speed, 20 % LED lighting and 22°C for 30 minutes

Day simulation: 60 % fan speed, 60 % LED lighting and 22°C for 12 hoursDay simulation: 60 % fan speed, 60 % LED lighting and 22°C for 12 hours

5

Night simulation: 0 % LED lighting and 22°C for 11 hours with unchanged fan speed of 20 %

Evening simulation: 20 % fan speed, 20 % LED lighting and 22°C for 30 minutesEvening simulation: 20 % fan speed, 20 % LED lighting and 22°C for 30 minutes

6

Morning simulation: 20 % LED lighting and 22°C for 30 minutes with unchanged fan speed of 20 %

Night simulation: 0 % LED lighting and 22°C for 11 hours with unchanged fan speed of 20 %Night simulation: 0 % LED lighting and 22°C for 11 hours with unchanged fan speed of 20 %

7

Jump back to 1 and repetition of the programme from 3 (31 times)

Morning simulation: 20 % LED lighting and 22°C for 30 minutes with unchanged fan speed of 20 %Morning simulation: 20 % LED lighting and 22°C for 30 minutes with unchanged fan speed of 20 %