The appliance may only be used in accordance with the applicable national regulations and

laws on keeping experimental animals and only be operated by persons who have the re-

quired qualifications for working with experimental animals.

This chapter merely describes the special functions of the HPPlife appliance. You

must read the operating manual for the constant climate chamber HPP (standard

model) in order to acquaint yourself with the basic operation and the operating

modes.

4.1 ControlCOCKPIT

In addition to the standard functions, you can set the speed of the exhaust air unit on the

back of the chamber in the ControlCOCKPIT of the HPPlife appliance.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

TIMER

End 14:45

44

h

:44

m

Holz trocknen
aufheizen
09:12h

Fr 20.10.2010 20:31

min

000°C

ALARM of °C

max

000°C

auto off

99K

-

+

0

12

%rh

°C

GRAPH

off

ON

ON

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

HUMIDITY

30

%rh

Set 30

%rh

22

.4

37

.0

TEMP

Set

°C

°C

LIGHT

%

100

ALARM

max

38

.5°C

auto

0.0

K

min

35

.5°C

+-

Manual Mode

12.Sept.2012 13:44

5

13

14

15

16

1

2

3

19

20

11 12

17

18

6 7

8

9

10

TIMER

30

m

04

h

End 13:30 23.11.

FAN

%

0

4

Fig. 9

ControlCOCKPIT of HPPlife appliances in operating mode

1 Activation key for temperature setpoint

adjustment

Activation key for temperature setpoint

Activation key for temperature setpoint

2 Setpoint and actual temperature display

adjustment

adjustment

3 Fan speed display

Setpoint and actual temperature display

Setpoint and actual temperature display

4 Activation key for fan speed setting

Fan speed display

Fan speed display

5 Changing to the menu mode (see operat-

Activation key for fan speed setting

Activation key for fan speed setting

ing manual constant climate chamber

Changing to the menu mode (see operat-

Changing to the menu mode (see operat-

HPP)

6 Activation key for timer setting

7 On/Off switch

Activation key for timer setting

Activation key for timer setting

8 Timer display

9 Humidity control display

Timer display

Timer display

10 Humidity control activation key

Humidity control display

Humidity control display

11 Turn control for setpoint adjustment

12 Confirmation key (accepts setting made

Turn control for setpoint adjustment

Turn control for setpoint adjustment

with the turn control)

Confirmation key (accepts setting made

Confirmation key (accepts setting made

13 Activation key for LED interior lighting

setting

14 Display LED interior lighting

setting

setting

15 Appliance state and programme display

Display LED interior lighting

Display LED interior lighting

16 Activation key for the appliance state

Appliance state and programme display

Appliance state and programme display

17 Activation key setting the temperature

Activation key for the appliance state

Activation key for the appliance state

and humidity monitoring

Activation key setting the temperature

Activation key setting the temperature

18 Monitoring display

and humidity monitoring

and humidity monitoring

19 Graphic representation

Monitoring display

Monitoring display

20 Activation key for graphic representation

Graphic representation

Graphic representation