2 connecting the appliance, 3 switching on the appliance, Connection to a central compressed air supply – Memmert HPP Life Constant climate chambers User Manual
Page 11: Putting into operation, Fig. 8 switch on appliance, Uf 110

D32802 | Date 10/2014
11
Putting into operation
3.2 Connecting the appliance
Connect the appliance to the energy and water supply as described in the operating manual
for the constant climate chamber HPP.
Connection to a central compressed air supply
Connection to a central compressed air supply
By doing so, the temperature-humidity working
range of the appliance can be extended. This
makes dehumidification of the air in the
appliance possible even at 10 ºC down to 10
% relative humidity (also see page 8).
With a suitable pressure hose, connect the
chamber to an external compressed air supply
(max. 5 bar) at the rear of the appliance (Fig.
7).
Only oil-free compressed air may be used.
If 10 % humidity is not reached at 10 ºC,
the compressed air is not, or not sufficiently,
pre-dehumidified. In this case, it is possible to
dehumidify the compressed air before it is fed
into the chamber via a maintenance unit con-
nected upstream, available from MEMMERT.
Several chambers can be connected to this
maintenance unit.
3.3 Switching on the appliance
Switch on the appliance by pressing the On/Off switch on
the front of the appliance ( Fig. 8 ).
After the first start-up, the appliance display is set to
English by default. You will find instructions as to how
to change the language in the operating instructions
of the constant climate chamber HPP.
Fig. 7
Compressed air connection
(max. 5 bar)
Fig. 7
Fig. 7
Compressed air connection
Compressed air connection
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
Fig. 8
Switch on appliance