P/N : 033-01591 Rev. 6
5.1.8
mVOLT
Function:
Sets (returns) the current mV calibration parameter
Summary:
Sets (returns) the current sensor mV calibration parameter. This may be the
parameter listed on the sensor, the manufacturers data sheet, or may have been
derived through the calibration procedure.
Example:
MVOLT 2.123
←
5.1.9
2PCAL
Function:
Runs the 2 point calibration procedure
Summary:
Begins the 2 point calibration procedure. Follow the prompts from the DI-1000
to complete the calibration as described in the section on calibration above.
Example
2PCAL 250
←
5.1.10
W
Function:
Displays the current weight
Summary:
Displays the current measured weight in the currently active units
Example
W
←
-277.162536
A
5.1.11
WC
Function:
Continuously outputs current weight
Summary:
Produces a continuous stream of output. Each line of the output represents the
load in the currently set units. You may stop the streaming load readings by
pressing the Enter key.
Example
WC
←