Leica VT1000 A
4. Setup the Instrument
43
Fig. 25
4.7.2 Assembling the rotating stage assembly
The rotating stage assembly allows for simple
position correction of the specimen surface
when the specimen is glued in place.
• To do so, first remove the specimen vice as-
sembly (34,
Fig. 16
) by following the reverse
order of steps as described in
Chap. 4.7
.
• Unscrew the knurled screw (38) completely.
You can now take the specimen clamp (40)
off of the base plate (39) (
Fig. 23
).
• Attach the rotating stage assembly (42) to
the base plate so that the pin (41) fits into the
bore (45) provided.
• Then, screw the knurled screw (38) into the
thread and tighten it, thus fastening the ro-
tating stage assembly (41) to the base plate
(39).
• Install it in the main specimen bath (23) as
described in
Chap. 4.7
(Fig. 25).
Fig. 23
38
39
40
39
42
41
45
Fig. 24
38
44
• The specimen to be sectioned (44) is glued
to the round specimen plate using the cy-
anoacrylate adhesive provided.
• To orient the specimen, slightly loosen the
knurled screw (43) on the right side, adjust to
the desired position and retighten the screw.
44
43