APPENDIX
X
Z306 Instruction
manual.docx
© Hermle Labortechnik GmbH
Tabelle 6: Radius correction
Rotor no.
Adapter/Tube‐
rack Order no.
Drawing. no.
Radius
(cm)
Correction
(cm)
Swing out rotor
220.72 V04
605.004
220.72.60.03
14,6
0
605.005
220.72.61.03
14,6
0,0
605.000/001
220.72.06.03
14,2
‐0,4
705.002
220.72.73.04
14,0
‐0,6
705.003
220.72.74.04
14,0
‐0,6
705.005
220.72.76.04
13,7
‐0,9
705.007
220.72.79.14
14,0
‐0,6
705.008
220.72.80.14
14,0
‐0,6
705.009
220.72.84.04
14,6
0
705.010
220.72.85.04
14,0
‐0,6
705.012
220.72.62
14,0
‐0,6
705.013
220.72.63
13,9
‐0,7
705.014
220.72.64
13,1
‐1,5
705.015
220.72.65
14,0
‐0,6
705.016
220.72.66
14,0
‐0,6
220.72.88
13,9
‐0,7
Angle rotor
220.87 V09
220.87 V10
8,6
0
704.004
220.56.55.04
8,2
‐0,4
704.005
220.56.56.04
7,5
‐1,1
Angle rotor
220.96 V04
601.005
220.96.05.04
10,5
0
Special
220.96.17
9,8
‐0,6
220.96.17
With
220.96.08.04
9,0
‐1,5
220.96.12.04
10,7
220.96.12.04
With
9,8
‐0,6
220.96.08.04
220.96.05.04
With
9,7
‐0,8
220.96.08.04
701.011
220.96.06.04
9,7
‐0,8
701.012
220.96.07.04
8,3
‐2,2
Angle rotor
220.97 V04
603.000
220.97.04.04
9,8
0
Swing out rotor
220.50 V08
10,2
0