Labnet Z306 Hermle Universal Centrifuge User Manual
Page 14
![background image](/manuals/565100/14/background.png)
PRODUCT DESCRIPTION
8
Z306 Instruction
manual.docx
© Hermle Labortechnik GmbH
1.7 Basic adjustments
At commissioning of the centrifuge, you have the options to make the following basic settings:
- Acustic signal turn on / off
- Keyboard sound turn on -/ off
- Volume pre-selection of sound signal
- Song selection of sound signal „end of run“
1.7.1 Access to mode „Operating Data“
If the centrifuge is still turned off, press simultaneously the keys „time“ (6) and „lid“ (7) and turn on
the main switch of the centrifuge. Now release both keys again. As a result a display test is
excecuted for approx. 5 seconds. All possible indications will appear at the same time (see Figure
2).
Figure 2
Attention:
- Please notice that you must enter the program as described under point 1.7.1 to
change the adjustments of the points 1.7.2 - 1.7.6. After you have stored the
settings you change the normal program mode again by switch off the centrifuge
for a short while.
- All changed settings must be confirmed by the key „start“ (9). As an optical
confirmation appears the word „store“ in the display „rpm/rcf“(A-1) - Only then
the pre-selections are valid! (see figure 3)