beautypg.com

Labnet Z400 High Capacity Centrifuge User Manual

Page 21

background image

Error no.: 70

Error no.: 70

Error no.: 70

Error no.: 70

Error no.: 70

Interface of the frequency converter

Interface of the frequency converter

Interface of the frequency converter

Interface of the frequency converter

Interface of the frequency converter

Reason:

Communication of controller, power board, interface cable

and frequency converter is not working

Action:

Call service

Error no.: 82-83

Error no.: 82-83

Error no.: 82-83

Error no.: 82-83

Error no.: 82-83 Cutoff power board - frequency converter

Cutoff power board - frequency converter

Cutoff power board - frequency converter

Cutoff power board - frequency converter

Cutoff power board - frequency converter

Reason:

Overcurrent or undervoltage due to power supply fluctuations

Action:

Restart the unit, take care that the power supply is stable

Error no.: 84

Error no.: 84

Error no.: 84

Error no.: 84

Error no.: 84

Overtemperature at the driving

Overtemperature at the driving

Overtemperature at the driving

Overtemperature at the driving

Overtemperature at the driving

Reason:

Temperature at the converter or motor too high

Action:

Switch off the centrifuge. Wait for about 15 min. and switch

the unit on again

Error no.: 85-87

Error no.: 85-87

Error no.: 85-87

Error no.: 85-87

Error no.: 85-87 Failures

Failures

Failures

Failures

Failures

Reason:

Internal defect

Action:

Call service

Error no.: 90

Error no.: 90

Error no.: 90

Error no.: 90

Error no.: 90

Emergency lid release

Emergency lid release

Emergency lid release

Emergency lid release

Emergency lid release

Reason:

The centrifuge lid has been opened by the emergency lid release

during the run

Action:

Close centrifuge lid. Danger of accident!

Reason:

Control switch of the lid lock is defective

Action:

Call service

Error no.: 94

Error no.: 94

Error no.: 94

Error no.: 94

Error no.: 94

Voltage loss during run

Voltage loss during run

Voltage loss during run

Voltage loss during run

Voltage loss during run

Reason:

The power supply is below the tolerance for a short moment

Action:

Wait till stillstand of the rotor. Open centrifuge lid after the yellow

LED "lid" is lighting. Switch off and on the main switch.

This manual is related to the following products: