Labnet Z400 High Capacity Centrifuge User Manual
Page 18
8.2
Check list / Trouble shooting
The error message will be indicated by a certain number on the digital speed
display. At the same time "ERROR" appears on the preset display.
There is a distinction between two different kinds of errors. The digits on the
indication "speed" have the following meaning:
Error no. 1 - 49 (Forced stop)
If one of those errors occurs, the rotor will be braked from the preset speed to 0.
As soon as the rotor has stopped, the error message can be reset by opening and
closing the centrifuge lid.
Error no. 50 - 99 (Emergency stop)
If this occurs, the frequency converter will be switched off. This means that the
rotor will be stopped brakeless. To reset the error message you have to plug off
and plug in the power cord.
If the unit stops due to an error indication you should restart the unit to check if the
error occurs again.
The error numbers which are not listed in this chapter are not in use at the time of
publication and they are reserved for future use in widening the error recognition
program.
Error no.: 1
Error no.: 1
Error no.: 1
Error no.: 1
Error no.: 1
Imbalance
Imbalance
Imbalance
Imbalance
Imbalance
Reason:
Incorrect loading of the rotor
Action:
Balance your samples
Reason:
Incorrect adjustment of the imbalance switch
Action:
Imbalance switch has to be readjusted (call service)
Error no.: 2
Error no.: 2
Error no.: 2
Error no.: 2
Error no.: 2
Permanent imbalance signal
Permanent imbalance signal
Permanent imbalance signal
Permanent imbalance signal
Permanent imbalance signal
Reason:
Position of the imbalance switch not correct
Action:
Imbalance switch has to be readjusted (call service)
Reason:
Imbalance switch is defective
Action:
Imbalance switch has to be replaced (call service)